Q: Is there a code for blanket warmers, towel warmers and such equipment that is used in a medical setting? Since they do have a heating element would it be the same as having a space heater?

A: No… Portable space heaters are regulated by section 19.7.8 of the 2012 LSC. Blanket warmers are not regulated by the LSC. However, blanket warmers are unofficially regulated by the accreditation organization and by CMS. While there are not specific standards that address blanket warmers, surveyors will expect the hospital to regulate their blanket warmers via their own policies.

The hospital is expected to have a policy that sets the maximum temperature that the blankets may be. Usually, the maximum temp is set at 130 degrees F. Anything over that will likely be scrutinized by a surveyor. Blanket warmers are considered medical equipment and must be included in the medical equipment inventory, and they must be maintained in accordance with the manufacturer’s recommendations. 

Brad Keyes, CHSP, is the owner of KEYES Life Safety Compliance, and his expertise is in the management of the Life Safety Program, including the Environment of Care and Emergency Management programs.

 

 



November 14, 2018


Topic Area: Regulations, Codes & Standards


Recent Posts

Legionella Detected at Michigan Health Facility

Case follows a similar finding of Legionnaires’ disease in a former patient.


Crystal Spring Tower Opens at Carilion Roanoke Memorial Hospital

The new cardiovascular institute improves access and care across the region.


Hospital Evacuation Highlights Importance of Emergency Preparedness

Berger Hospital was evacuated after smoke was discovered in the basement.


Building a Culture of Infection Prevention

Investing in infection prevention can benefit healthcare organizations and patient outcomes.


Ground Broken on Sanford Health Fargo's Peltier Lodge

The 28,400-square-foot facility will feature spaces to support patients and loved ones during treatment and throughout their outpatient care and monitoring.


 
 


FREE Newsletter Signup Form

News & Updates | Webcast Alerts
Building Technologies | & More!

 
 
 


All fields are required. This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.