Focus: Fire-Life Safety / Column

Regulations, Codes & Standards Q&A: Receptacle testing

Brad Keyes discusses regulations for receptacle testing

By Brad Keyes / Special to Healthcare Facilities Today


Q: One of our communities in Minnesota has just completed their Life Safety Survey and during the brief exit the Fire Marshall indicated that a yearly test on all electrical outlets in each resident room needs to be documented and tested on an annual basis. Could you please point me to the Life Safety Code or the Health Care Facilities Code that pertains to this testing and documentation? 

A: According to section 6.3.4.1.3 of NFPA 99-2012, receptacles not listed as hospital-grade, at patient bed locations and in locations where deep sedation or general anesthesia is administered, shall be tested at intervals not exceeding 12 months. 

But sections 6.3.4.1.1 and 6.3.4.1.2 of NFPA 99-2012 do say that hospital-grade receptacles are required to be tested when they are initially installed, but additional testing of the receptacles in patient care rooms shall be performed at intervals defined by the documented performance data, which would be either the manufacturer’s recommendations, or historical evidence. 

So, to summarize: You do have to test non-hospital-grade receptacles on an annual basis in patient care rooms and anesthetizing locations. But hospital-grade receptacles are not required to be tested annually, but the frequency of the test is based on the performance data.

Brad Keyes, CHSP, is the owner of KEYES Life Safety Compliance, and his expertise is in the management of the Life Safety Program, including the Environment of Care and Emergency Management programs.

 



December 26, 2019


Topic Area: Regulations, Codes & Standards


Recent Posts

Mature Dry Surface Biofilm Presents a Problem for Candida Auris

Multiple methods are described in the literature, but no consensus has been reached for disinfection efficacy tests against biofilms.


Sutter Health's Arden Care Center Officially Opens

With an adaptive reuse of an underutilized office building, the 70,000 square-foot facility was renovated to meet current healthcare standards.


Insight Hospital and Medical Center Falls to Data Breach

The investigation determined that an unauthorized individual accessed the network between August 22, 2025, and September 11, 2025.


The High Cost of Healthcare Violence

As workplace violence increases, healthcare facilities face mounting financial and operational disruptions- prompting legislative action.


EVS Teams Can Improve Patient Experience in Emergency Departments

A report confirmed that cleanliness of the ED was the third most impactful element on patient experience surveys.


 
 


FREE Newsletter Signup Form

News & Updates | Webcast Alerts
Building Technologies | & More!

 
 
 


All fields are required. This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.